6.4 CONDUCTIVITY
• From GLP data main menu select the “Conduc-
tivity” option and press OK. This menu allows
to view data for conductivity, actual conductiv-
ity and salinity calibrations.
All the information about the last conductivity calibration appears: calibrated
point, cell constant value, calibration type (conductivity, actual conductivity or
salinity), time and date.
Use the arrow keys to scroll the last 5 memorized calibrations.
For conductivity calibration GLP data the following screens are available: con-
ductivity, actual conductivity, salinity.
NONO
NONO
NO
TETE
TETE
TE The C letter near the conductivity calibration indicates a custom point,
while the H means HANNA standard value.
If no conductivity calibration has been performed, a warning message
advise the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.
6.5 ATMOSPHERIC PRESSURE
From GLP data menu select the “Atmospheric pressure” option and press OK.
All the information about the last atmospheric pressure calibration appears:
custom calibration point, time and date.
• In case of restoring factory calibration data, the
display will show the warning message “Fac-
tory calibr. restored”.
Use the arrow keys to scroll the last 5 memo-
rized calibrations.
NONO
NONO
NO
TETE
TETE
TE If no atmospheric pressure calibration has been performed, a warning
message advise the user. Press OK to return to the previous screen.
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.