Press CONFIRM to accept the value or BUFFER to select another buffer using
the arrow keys.
After the first calibration point is confirmed, immerse the probe in the second
buffer solution and stir gently.
When the reading is stable and close to the selected buffer, the display shows
the message “Ready”.
Press CONFIRM to accept the value or BUFFER to change the buffer.
After the second calibration point is confirmed, proceed by immersing the probe
in the third buffer solution, stirring gently and waiting for stable reading.
When the calibration is completed, the display shows the following messages:
“Storing data on probe, please wait...”, “Updating GLP data, please wait ...”
and “Calibration completed”.
Press OK to return to the calibration menu.
To return to the main menu, press ESC repeatedly.
NONO
NONO
NO
TETE
TETE
TE The user can terminate the pH calibration mode at any time, by press-
ing the ESC key.
Custom buffer calibration
HI 9828 allows a single point procedure to cali-
brate with a custom buffer value.
Select this option by pressing BUFFER and then
CUSTOM keys while the meter is waiting for
stable reading.
The display shows a window with a text box for inserting the desired custom
value. The valid range for custom buffer is 0.00 to 14.00 pH.