4.3 DISSOLVED OXYGEN CALIBRATION
If D.O.% saturation range is calibrated, D.O. concentration range will also be
calibrated, and vice-versa.
The D.O.% saturation value is referred to the D.O. concentration in air (100%).
For this reason it is recommended to calibrate the probe near the area where the
measurements will be taken.
Also note that the D.O. concentration values are based on D.O.% saturation,
temperature, salinity and atmospheric pressure. It is recommended to use a stan-
dard solution or a reference D.O. meter to compare readings during calibration.
The calibration of D.O.% saturation range can be performed at 1 or 2 standard
points (0% and 100%), or at a single custom point (50 to 500%).
The calibration of D.O. concentration range can be performed at a single cus-
tom point (4 to 50 mg/L).
4.3.1 Procedure
Choose the D.O. calibration mode in the main
calibration menu, then select the calibration type
using the arrow keys and confirm by pressing OK.
D.O. % saturation
This calibration starts by default with 100 %.
Fill the calibration beaker with approximately 4
mm (5/32”) of distilled water and screw it on
the probe.
The message NOT READY is displayed until a
stable reading is reached.
To change the standard calibration point, press
the CAL. POINT softkey and select the desired
standard point.
• To insert a different calibration value, press CAL.
POINT and then CUSTOM. Insert the desired
value using the keyboard.
When the reading is stable, READY and CONFIRM appear. Press CONFIRM to
store the calibration point.
After the first calibration point is confirmed, put the probe in a zero oxygen
standard solution and wait for stable reading.