LAST RELATIVE mV CALIBRATION DATA
Last Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press GLP key while in Relative mV
measurement mode.
The instrument will display the Relative mV GLP information: calibration date,
time and offset.
LAST ISE CALIBRATION DATA
Last ISE calibration data is stored automatically after a successful calibration.
To view the ISE calibration data, press GLP while in ISE measurement mode.
The instrument will display the ISE calibration information: calibration date,
time, slope, calibration status and electrode type.
Notes: Press GLP or ESC at any moment and the instrument will return
to measurement mode.
If calibration has not been performed, the instrument displays
“No user calibration” message.
The calibration standards from previous calibration are displayed
in video inverse mode.