Hanna-instruments HI 98180 Bedienungsanleitung Seite 19

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 72
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 18
19
After the third calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the fourth
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
Press CFM to confirm calibration.
The calibrated value and the
fifth
expected buffer will be displayed.
After the fourth calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the fifth buffer solution and
stir gently. The temperature probe should be close to the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, the CFM
functional key is displayed.
Seitenansicht 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 71 72

Kommentare zu diesen Handbüchern

Keine Kommentare