Hanna-instruments HI 98180 Bedienungsanleitung Seite 28

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 72
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 27
28
Press CAL. The primary LCD will displays the ion concentration in the selected
unit or “---” if not calibrated and first standard value.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the CFM
functional key is displayed.
Press CFM to confirm calibration.
The calibrated value and the second expected standard value will be
displayed.
After the first calibration point is confirmed, immerse the ISE electrode
approximately 4 cm (1½”) into the second calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the CFM
functional key is displayed.
Press CFM to confirm calibration.
The calibrated value and the third expected standard value will be
displayed.
After the second calibration point is confirmed, immerse the ISE electrode
approximately 4 cm (1½”) into the third calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the CFM
functional key is displayed.
Seitenansicht 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 71 72

Kommentare zu diesen Handbüchern

Keine Kommentare