Pay attention to push the cuvette completely down in the holder.
In order to avoid reagent leaking and to obtain more accurate measure-
ments, close the cuvette first with the supplied HDPE plastic stopper
and then the black cap.
Whenever the cuvette is placed into the measurement cell, it must be dry
outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
Shaking the cuvette can generate bubbles in the sample, causing higher
readings. To obtain accurate measurements, remove such bubbles by
swirling or by gently tapping the cuvette.
Do not let the reacted sample stand too long after reagent is added, or
accuracy will be lost.
It is possible to take multiple readings in a row, but it is recommended to
take a new zero reading for each sample and to use the same cuvette for
zeroing and measurement when possible (for most precise results follow the
measurement procedures carefully).
Discard the sample immediately after the reading is taken, or the glass might become permanently
stained.
All the reaction times reported in this manual are at 20°C (68°F). As a general rule of thumb, they
should be doubled at 10°C (50°F) and halved at 30°C (86°F).
REAGENT BLANK CORRECTION
Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly
in the same way, only the blank is deionized water instead of sample.
A blank cuvette may be used more than once: stability and storing conditions are described for each
method in the related chapter.
INTERFERENCES
In the method measurement section the most common interferences that may be present in an average
wastewater matrix have been reported. It may be that for a particular treatment process other
compounds do interfere with the method of analysis.
METHOD REFERENCE TABLE
HI83212 - POWER PLANT UTILITIES
Method Method Page
description
1 Ammonia MR 17
2 Ammonia LR 19
3 Free Chlorine 21
4 Total Chlorine 24
5 Copper HR 27
6 Copper LR 29
7 Hydrazine 31
Method Method Page
description
8 Molybdenum 33
9 Phosphate HR 36
10 Phosphate LR 38
11 Phosphorus 40
12 Silica 42
13 Silver 45
HEALTH & SAFETY
The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the
Material Safety Data Sheet (MSDS) before performing tests.
Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully.
Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.