Remove the cuvette.
Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawater
analysis). Replace the cap and mix the solution.
Add 4 drops of HI 93715B-0 Second Reagent (10 drops for seawater
analysis). Replace the cap and mix the solution.
Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds and press
READ. When the timer ends the meter will perform the reading. The
instrument displays the results in mg/L of ammonia nitrogen (NH
3
-N).
Press the s or t to access the second level of functions.
Press the Chem Frm functional key to convert the result in mg/L of ammonia (NH
3
) and ammonium (NH
4
+
).
Press the s or t to go back to the measurement screen.
INTERFERENCES
Interference may be caused by:
acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic chloramines, sulfide, various
aliphatic and aromatic amines.
Ammonia LR
SPECIFICATIONS
Range 0.00 to 3.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.04 mg/L ±4% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93700A-0 First Reagent 4 drops (6 drops for seawater)
HI 93700B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93700-01 Reagents for 100 tests
HI 93700-03 Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Select the
Ammonia LR
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the meter is
zeroed and ready for measurement.
AMMONIA LOW RANGE
Ammonia MR
x 4
x 4