Hanna-instruments HI 83212 Bedienungsanleitung Seite 19

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 27
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 18
36 37
Phosphate HR
Add the content of one packet of HI 93717B-0 Phosphate
HR Reagent B to the cuvette. Replace the cap and shake
gently until completely dissolved.
Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior
to the measurement or, alternatively, wait for 5 minutes
and press READ. When the timer ends the meter will
perform the reading. The instrument displays the results in
mg/L of phosphate (PO
4
3-
).
Press the s or t to access the second level of functions.
Press the Chem Frm functional key to convert the result in mg/L of phosphorus (P) and phosphorus
pentoxide (P
2
O
5
).
Press the s or t to go back to the measurement screen.
INTERFERENCES
Sulfide
Chloride above 150000 mg/L )
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L
Phosphate HR
PHOSPHATE HIGH RANGE
SPECIFICATIONS
Range 0.0 to 30.0 mg/L
Resolution 0.1 mg/L
Accuracy ±1 mg/L ±4% of reading
Typical EMC Dev. ±0.1 mg/L
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition,
Amino Acid method. The reaction between phosphate and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93717A-0 Molybdate 10 drops
HI 93717B-0 Reagent B 1 packet
REAGENT SETS
HI 93717-01 Reagents for 100 tests
HI 93717-03 Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Select the
Phosphate HR
method using the procedure described
in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the meter
is zeroed and ready for measurement.
Remove the cuvette.
Add 10 drops of HI 93717A-0 Molybdate reagent.
x 10
10 ml