1Instruction ManualHI 96713CPhosphate Low Range ISMwww.hannainst.com

Seite 2 - TABLE OF CONTENTS

10Over Range: A flashing value of the maximum concentrationindicates an over range condition. The concentration of thesample is beyond the programmed

Seite 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Seite 4 - ABBREVIATIONS

12MEASUREMENT PROCEDURE• After the reading it is important to discard immediately the sample, otherwise the glassmight become permanently stained.• Al

Seite 5 - PRECISION AND ACCURACY

13• Place the cuvette into the cuvette holderand ensure that the notch on the cap ispositioned securely into the groove.• Press ZERO/CFM and the lamp,

Seite 6 - PRINCIPLE OF OPERATION

14• Alternatively, wait for 3 minutes then justpress READ/ /TIMER. In all cases, thelamp, cuvette and detector icons willappear on the display, depend

Seite 7 - FUNCTIONAL DESCRIPTION

15VALIDATION PROCEDUREUse the validation procedure to ensure that the instrument is properly calibrated.Warning: Do not validate the instrument with a

Seite 8

16• Press CAL CHECK and the lamp, cuvetteand detector icons together with “CALCHECK” will appear on the display,depending on the measurement phase.• A

Seite 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Seite 10

18• After a few seconds the display willshow “-0.0-”. The meter is now zeroedand ready for calibration. The blinking“READ” asks for reading calibratio

Seite 11

19• Press GLP/ to edit the desired year(2009-2099). If the key is kept pressed,the year number is automaticallyincreased.• When the correct year has

Seite 12 - MEASUREMENT PROCEDURE

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DE

Seite 13

20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco

Seite 14

21• Press READ/ /TIMER to enter in thefactory calibration restore screen. Theinstrument asks for confirmation of usercalibration delete.• Press ZERO/C

Seite 15 - VALIDATION PROCEDURE

22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract

Seite 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Seite 17 - CALIBRATION PROCEDURE

24MAN96713 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Seite 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Seite 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96713 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Seite 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ± 4% of reading @ 25°CTypical EMC Deviation ±0.01

Seite 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Seite 22 - ACCESORIES

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette ho

Seite 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to me

Seite 24 - (ROMANIA)

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are alsodisplayed when the obtained values are outs