Seite 1 - Free and Total Chlorine

1www.hannainst.comInstruction ManualHI 96711CFree and Total Chlorine ISM

Seite 2 - TABLE OF CONTENTS

18 19The reading should be within specifications as reported in the CAL CHECKTM StandardCertificate. If the value is found out of the specifications,

Seite 3 - PRELIMINARY EXAMINATION

20 21• Remove the cuvette.• Place the specific CAL CHECKTM StandardCuvette B into the cuvette holder, for:Free Chlorine: B, HI 96701-11orTotal Chlorin

Seite 4 - PRECISION AND ACCURACY

22 23To display the calibration date:• Press and hold for three secondsRANGE/GLP/ to enter GLP mode. Thecalibration month and day will appear onthe m

Seite 5 - FUNCTIONAL DESCRIPTION

24 25ACCESORIESREAGENT SETHI 93701-01 Reagents for 100 Free Chlorine testsHI 93701-03 Reagents for 300 Free Chlorine testsHI 93711-01 Reagents for 100

Seite 6 - ERRORS AND WARNINGS

26 27Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environ

Seite 7

28MAN96711 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Seite 8 - RANGE SELECTION

2 3TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 4GENERAL

Seite 9 - MEASUREMENT PROCEDURE

4 5GENERAL DESCRIPTIONABBREVIATIONSThe HI 96711 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a

Seite 10 - VALIDATION PROCEDURE

6 7PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cro

Seite 11 - CALIBRATION PROCEDURE

8 91) RANGE/GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cu

Seite 12

10 11ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are

Seite 13 - BATTERY REPLACEMENT

12 13The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts ma

Seite 14 - WARRANTY

14 15• Press RANGE/GLP/ to change the range.The range can be changed any time when theinstrument is in measurement mode.The selected range is memoriz

Seite 15 - MAN96711 12/10

16 17• After a few seconds, the display willshow “-0.0-”. The meter is now zeroedand ready for measurement.• Remove the cuvette.• Add the content of o