Hanna-instruments HI 84500 Bedienungsanleitung Seite 23

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 36
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 22
23
At the end of the titration the instrument displays the concentration in ppm of SO
2
. The titration
curve can be viewed by pressing Plot ON. Press Plot OFF to exit this mode.
Press LOG to record the concentration value and the titration curve into the instrument’s
memory. A message will be displayed for a few seconds indicating the amount of free log
space. Up to 200 log samples can be recorded in the instrument’s memory.
Press Restart to begin a new titration or ESC to return to the titration menu.
If the concentration exceeds the range limits (>40.0 ppm for Low Range, >400 ppm for
High Range), the exceeded range limit will be displayed blinking. Another titration can be
started by pressing Restart.
“Wrong input” error message appears when the input reading (mV) exceeds the specified
limits. The mV value and the concentration will blink indicating an error.
This screen appears when the stirrer is not working properly. Check the stir bar and beaker
content. Press Restart to try again.
Seitenansicht 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 35 36

Kommentare zu diesen Handbüchern

Keine Kommentare