MEASUREMENT PROCEDURE
Refer to “Setup Menu” (see page 10) to set up the instrument for your measurement.
For best accuracy, before taking any measurement, ensure that the pump is calibrated on the
selected range following the “Pump Calibration Procedure” (see page 19).
Select the corresponding range according to the table below:
Ensure the pump is primed with the correct titrant for the selected range (HI 84500-50 Low
Range Titrant or HI 84500-51 High Range Titrant).
Sample preparation: Use a clean pipette to precisely add the appropriate amount of wine
sample to a clean 100 mL beaker as indicated below:
egnaRwoL
)elpmasLm05(
egnaRhgiH
)elpmasLm05(
mpp0.04ot0.1 mpp004ot03
Low Range (Free & Total SO
2
) - 50 mL
High Range (Free & Total SO
2
) - 50 mL
Note: The volume of wine added is critical to the measurement
accuracy. Pipettes are recommended.
Failure to use a clean pipette will result in erroneous
readings.
Fill the 20 mL beaker up to the 5 mL mark with the HI 84500-60
Acid Reagent and add the contents to the 100 mL beaker.
Add the contents of one packet of HI 84500-62 Stabilizer
Packet to the 100 mL beaker.
Press Titrator.
This screen appears when the stirrer is not
working properly. Check the stir bar and beaker
content. Press Restart to try again.
This error message appears when the pump is
not working properly. Check the tubing, valve
and syringe. Press Restart to try again.