the calibration time, yymmddhhmmss (12 chars).
electrode condition, with sign (3 chars). The
“-01” code means not calculated.
Rel mV calibration data (if available), which contains:
the calibration offset, with sign (7 chars)
the calibration time, yymmddhhmmss (12 chars).
ISE calibration data (if available), which contains:
the number of calibrated standards (1 char)
the ion charge, with sign (0 means undefined)
(2 chars)
the calibration slope, with sign and decimal point
(7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each standard)
type (1 char): 0 - always standard solution.
status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration.
warnings during calibration (2 chars): 00 -
no warning.
standard value, with sign and decimal point
(7 chars).
calibration time, yymmddhhmmss (12 chars).
PAR Requests the setup parameters setting.
The answer string contains:
Instrument ID (4 chars)
Calibration alarm time out for pH range (2 chars)
Calibration alarm time out for ISE range (2 chars)
SETUP information (2 chars): 8 bit hexadecimal
encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
0x08 - Offset calibration (else Point calibration)
The number of custom buffers (1 char)
The custom buffer values, with sign and decimal
point, for each defined custom buffer (7 chars)
Ion charge (2 chars)
NSLx Requests the number of logged samples (4 chars).
Command Parameter (1 char):
x - range (see Note)