Hanna-instruments HI 2216 Bedienungsanleitung Seite 36

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 52
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 35
36
CHR xx Change the instrument range according with the param-
eter value (xx):
xx=00 pH range/0.001 resolution
xx=01 pH range/0.01 resolution
xx=02 pH range/0.1 resolution
xx=03 mV range
xx=04 Relative mV range
xx=05 ISE range
The instrument will answer for these commands with:
<STX> <answer
>
<ETX>
where: <STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)
COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:
<STX> <answer> <checksum> <ETX>
where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.
All the answer messages are with ASCII characters.
RAS Causes the instrument to send a complete set of readings in
according with the current range:
pH, temperature and mV reading on pH range.
Rel mV, absolute mV and temperature reading on
Relative mV range.
ISE, mV and temperature reading on ISE range.
The answer string contains:
Meter mode (2 chars):
00 - pH range (0.001 resolution)
01 - pH range (0.01 resolution)
02 - pH range (0.1 resolution)
03 - mV range
04 - Rel mV range
05 - ISE range
Seitenansicht 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 ... 51 52

Kommentare zu diesen Handbüchern

Keine Kommentare