Instrument ID (4 chars)
Calibration alarm time out for pH range (2 chars)
SETUP information (2 chars): 8 bit hexadecimal
encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
0x08 - Offset calibration (else Point calibration)
The number of custom buffers (1 char)
The custom buffer values, with sign and decimal
point, for each defined custom buffer (7 chars)
NSLx Requests the number of logged samples (4 chars).
Command Parameter (1 char):
x - range (see Note)
LODPxxx Requests the xxx
th
pH record logged data.
LODMxxx Requests the xxx
th
mV/Rel mV record logged data.
LODPALLff Requests all pH Log on demand - ff=frame number
LODMALLff Requests all mV/Rel mV Log on demand - ff=frame number
Note:The log on demand can be downloaded in frames of 8
records (01 is first frame number)
LLSxff: Requests information about all lots in frames of 10 lots (a frame
contains information about 10 lots) (HI 2215 only)
Command Parameters:
x - range (see Note)
ff - requested frame number - first frame is labeled 01
GLDxxxff: Requests the records of the “xxx” lot number (HI 2215 only).
The records are sent in frames of 10 records; “ff” is the frame
number (01 first frame). (Example: Lot 13 has 53 records.
The records will be sent in 6 frame, 5 with10 records and 1
with 3 records.)
Command Parameters:
xxx - Lot number (eq: for lot number 1 xxx = 001)
ff - requested frame number - first frame is labeled 01
Errors: “Err3” is sent if the Log on demand is empty.
“Err4” is sent if the requested set parameter is not available.
“Err5” is sent if an argument of the command is not correct.
“Err6” is sent if the requested range is not available.
“Err7” is sent if the instrument is in logging mode.
“Err8” is sent if the instrument is not in measurement mode.
Invalid commands will be ignored.
Note: P - request for pH range.
M - request for mV and Rel mV ranges.