29
INSTRUMENT ID
Press CAL when the “InId” is displayed. The instrument ID (“0000“ to
“9999“) and “CFM” will start blinking.
Press the ARROW
keys to change the instrument ID value.
Press CFM to save the modified instrument ID value.
Press CAL to escape without saving.
Note: The instrument ID is downloaded to a PC as part of a logged
data, set to identify its origin.
TEMPERATURE UNIT
Press CAL when “tnP“ is displayed. The temperature unit and “CFM”
will start blinking.
Press the ARROW keys to change the option.
Press CFM to save the modified temperature unit.
Press CAL to escape without saving.
Press the ARROW
keys to change the beep status (On or OFF).
Press CFM to save the modified beep status.
Press CAL to escape without saving.
When enable, beep sounds as a short beep every time a key is pressed or
when the calibration can be confirmed.
A long beep alert that the pressed key is not active or a wrong condition
is detected while in calibration.