Hanna-instruments HI 83266 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messgeräte Hanna-instruments HI 83266 herunter. Hanna Instruments HI 83266 User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - HI 83226

1Instruction ManualHI 83226Multiparameter BenchPhotometerfor Pool & Spa Applicationswww.hannainst.com

Seite 2 - GENERAL DESCRIPTION

18 19HI 83226 offers an interactive contextual help mode that assists the user at any time.To access the help screens press HELP.The instrument will d

Seite 3

20 21SPECIFICATIONSRange 0 to 500 mg/L (as CaCO3)Resolution 5 mg/LAccuracy ±5 mg/L ±10% of reading at 25 °CTypical EMC ±5 mg/LDeviationLight Source Tu

Seite 4 - WATER INDEX REFERENCE TABLES

22 23BromineBROMINESPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.08 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviatio

Seite 5 - PRECISION AND ACCURACY

24 25Calcium HardnessCALCIUM HARDNESSSPECIFICATIONSRange 0 to 500 mg/L (as CaCO3)Resolution 5 mg/LAccuracy ±10 mg/L ±5% of reading at 25 °CTypical EMC

Seite 6 - FUNCTIONAL DESCRIPTION

26 27Calcium HradnessFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/L from 0

Seite 7 - USING CUVETTES

28 29• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.• Add the content of on

Seite 8 - OPERATIONAL GUIDE

30 31Total Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.• Add 1 p

Seite 9 - SETUPSETUP

32 33• Remove the cuvette.• Add the content of one packet of HI 93702-0Copper Reagent. Replace the cap and shake gentlyfor 15 seconds.• Reinsert the c

Seite 10 - HELP MODE

34 35• Remove the cuvette.• Add the content of one packet of HI 93702-0Copper Reagent. Replace the cap and shake gentlyfor 15 seconds.• Add the conten

Seite 11 - ALKALINITY

36 37• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

Seite 12 - INTERFERENCES

2 3Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will pr

Seite 13 - CALCIUM HARDNESS

38 39Iron• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively,

Seite 14 - FREE CHLORINE

40 41Ozone• Press the Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement.• Remove the cuvette.• Add the content of

Seite 15 - TOTAL CHLORINE

42 43pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pH at 25 °CTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow

Seite 16 - FREE COPPER

44 45The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messag

Seite 17 - TOTAL COPPER

46 47ACCESSORIESREAGENT SETSHI 93701-01 100 free chlorine tests (powder)HI 93701-03 300 free chlorine tests (powder)HI 93701-F 300 free chlorine tests

Seite 18 - CYANURIC ACID

48Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Seite 19

4 5A major family leisure pursuit is the enjoyment of Swimming Pool and Spa facilities world-wide. A basic necessityof Pool water treatment, to ensure

Seite 20

6 7AlkalinityAlkalinity is the measure of the total concentration of alkaline substances, mostly bicarbonates, dissolved in thewater. The higher the a

Seite 21

8 9SPECIFICATIONSLight Life Life of the instrumentLight Detector Silicon PhotocellEnvironment 0 to 50°C (32 to 122°F);max 90% RH non-condensingPower S

Seite 22

10 11INSTRUMENT DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD)5) Splash proof keypad6) ON/OFF power swi

Seite 23 - DATA MANAGEMENT

12 13(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hangingon the tip of the syringe, if so eliminate th

Seite 24 - HANNA LITERATURE

14 15METHOD REFERENCE TABLEMethod Method Pagedescription1 Alkalinity 20 2 Bromine 223 Calcium Hardness 244 Free Chlorine 275 Total Chlorine 296 F

Seite 25 - MAN83226 06/11

16 17DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measu

Kommentare zu diesen Handbüchern

Keine Kommentare