1Instruction ManualHI 83226Multiparameter BenchPhotometerfor Pool & Spa Applicationswww.hannainst.com
18 19HI 83226 offers an interactive contextual help mode that assists the user at any time.To access the help screens press HELP.The instrument will d
20 21SPECIFICATIONSRange 0 to 500 mg/L (as CaCO3)Resolution 5 mg/LAccuracy ±5 mg/L ±10% of reading at 25 °CTypical EMC ±5 mg/LDeviationLight Source Tu
22 23BromineBROMINESPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.08 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviatio
24 25Calcium HardnessCALCIUM HARDNESSSPECIFICATIONSRange 0 to 500 mg/L (as CaCO3)Resolution 5 mg/LAccuracy ±10 mg/L ±5% of reading at 25 °CTypical EMC
26 27Calcium HradnessFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/L from 0
28 29• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.• Add the content of on
30 31Total Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.• Add 1 p
32 33• Remove the cuvette.• Add the content of one packet of HI 93702-0Copper Reagent. Replace the cap and shake gentlyfor 15 seconds.• Reinsert the c
34 35• Remove the cuvette.• Add the content of one packet of HI 93702-0Copper Reagent. Replace the cap and shake gentlyfor 15 seconds.• Add the conten
36 37• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
2 3Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will pr
38 39Iron• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively,
40 41Ozone• Press the Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement.• Remove the cuvette.• Add the content of
42 43pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pH at 25 °CTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow
44 45The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messag
46 47ACCESSORIESREAGENT SETSHI 93701-01 100 free chlorine tests (powder)HI 93701-03 300 free chlorine tests (powder)HI 93701-F 300 free chlorine tests
48Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe
4 5A major family leisure pursuit is the enjoyment of Swimming Pool and Spa facilities world-wide. A basic necessityof Pool water treatment, to ensure
6 7AlkalinityAlkalinity is the measure of the total concentration of alkaline substances, mostly bicarbonates, dissolved in thewater. The higher the a
8 9SPECIFICATIONSLight Life Life of the instrumentLight Detector Silicon PhotocellEnvironment 0 to 50°C (32 to 122°F);max 90% RH non-condensingPower S
10 11INSTRUMENT DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD)5) Splash proof keypad6) ON/OFF power swi
12 13(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hangingon the tip of the syringe, if so eliminate th
14 15METHOD REFERENCE TABLEMethod Method Pagedescription1 Alkalinity 20 2 Bromine 223 Calcium Hardness 244 Free Chlorine 275 Total Chlorine 296 F
16 17DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measu
Kommentare zu diesen Handbüchern