SPECIFICATIONS
Range 0 to 500 mg/L (as CaCO
3
)
Resolution 5 mg/L
Accuracy ±5 mg/L ±10% of reading at 25 °C
Typical EMC ±5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Colorimetric Method. At different alkalinity levels a distinctive range of colors from yellow
to green and greenish blue will develop.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93755-0 Alkalinity Indicator Reagent 1 mL
REAGENT SETS
HI 93755-01 Reagents for 100 tests
HI 93755-03 Reagents for 300 tests
For other accessories see page 46.
MEASUREMENT PROCEDURE
Select the
Alkalinity
method using the procedure described in the
Method Selection
section (see page 15).
Fill the cuvette with 10 mL of unreacted sample, up to the mark,
and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
ALKALINITY
Alkalinity
10 mL
Remove the cuvette.
Carefully add exactly 1 mL of HI 93755-0 Liquid Alkalinity
Reagent using the supplied syringe.
Replace the cap and invert 5 times.
Reinsert the cuvette into the instrument and close the lid.
Press Read to start the reading.
The instrument displays the results in mg/L of alkalinity (CaCO
3
).
Note: If using a meter with software version 1.14 or earlier, readings can be improved for samples
with less than 75 ppm alkalinity by adding 0.7 mL of reagent instead of 1.0 mL.