HI 8424NEW has been accurately precalibrated for temperature at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for tempera-
ture at least once a year.
Contact your dealer or the nearest HANNA instruments
®
Customer Service
Center for more information.
HI 8424NEW has been accurately precalibrated for mV range at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for mV
readings at least once a year.
Contact your dealer or the nearest HANNA instruments
®
Customer Service
Center for more information.
mm
mm
m
V CALIBRATIONV CALIBRATION
V CALIBRATIONV CALIBRATION
V CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATIONTEMPERATURE CALIBRATION
TEMPERATURE CALIBRATION
16
TROUBLESHOOTING
Symptom Problem Solution
Slow reponse or Dirty pH electrode Soak the electrode tip
excessive drift in HI 7061 solution for
30 minutes
Reading fluctuates Clogged/dirty junction Soak the electrode tip
up and down (noise) or low electrolyte level in warm HI 7082
(refillable electrodes) solution for one hour,
then rinse it with
distilled water
(refill with fresh
electrolyte if necessary)
Blinking full scale Reading is out of range
value
Blinking "°C" (or "°F") Temperature probe is not connected or broken
"WRONG" & Calibration error Check buffer solution or
blinking dashes replace pH electrode
Blinking battery Low battery level Replace battery
symbol
Meter shuts off Auto-off enabled or
dead battery Replace battery
"Clr" message Loaded default Perform pH calibration
pH calibration values
"Er1" and Er2" EPROM error Contact your dealer or
messages HANNA Service Center
Note: For field applications, it is always recommended to keep a conditioned spare elec-
trode handy. When anomalies cannot be resolved with simple maintenance, change
the electrode and recalibrate the meter.