At the end of the measurement, the third
calibration point (100 NTU) is displayed on
the primary LCD, “CAL P.3” on the secondary
LCD, and “READ” tag will blink.
At this moment it is possible to exit calibration
by pressing CAL. The instrument will memorize
the two-point calibration data and will return
to measurement mode.
THREE-POINT CALIBRATION
To perform a three-point calibration, continue the
procedure with the following steps:
Remove the second standard cuvette.
Place the 100 NTU standard cuvette (or the
prepared formazin standard) into the holder,
with the cuvette mark aligned with the sign on
the instrument.
Close the lid and press READ u. The display
will show blinking dashes and the icons for
cuvette, detectors and lamp will appear during
measurement.
At the end of the measurement, the fourth
calibration point (750 NTU) is displayed on
the primary LCD and “CAL P.4” on the
secondary LCD.
At this moment it is possible to exit calibration
by pressing CAL. The instrument will memorize
the three-point calibration data and will return
to measurement mode.
FOUR-POINT CALIBRATION
To perform a four-point calibration, continue the
procedure with the following steps:
Remove the third standard cuvette.
Place the 750 NTU standard cuvette (or the
prepared formazin standard) into the holder,
with the cuvette mark aligned with the sign on
the instrument.