Hanna-instruments HI 98202 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messgeräte Hanna-instruments HI 98202 herunter. Hanna Instruments HI 98202 User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 2
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
pNa
HI 98202
Soft Water Tester
GENERAL INFORMATION
The pNa tester utilizes a sodium ion-selective
electrode to determine the activity of free sodium
in solution (pNa = -log a
Na
).
In dilute solutions, the activity coefficient is nearly
1 and in such solutions pNa is a good indicator
of the sodium ion concentration. The relationship
between the pNa scale and the g/L Na
+
scale is
explained in the chart below.
A double junction reference is used to ensure a
highly stable reading. If you suspect that the
calibration has drifted, you can recalibrate the
meter by using a solution of known concentration.
Adjust the reading with the calibration trimmer.
HOW TO USE THE CHART
• Locate the pNa reading on the horizontal axis
of the chart.
Move vertically upwards to intersect with the
45º line.
• Move horizontally and read the g/L Na value.
E.g. pNa = 1.4
g/L = 0.9
www.hannainst.com
Seitenansicht 0
1 2

Inhaltsverzeichnis

Seite 1

pNaHI 98202Soft Water TesterGENERAL INFORMATIONThe pNa tester utilizes a sodium ion-selectiveelectrode to determine the activity of free sodiumin sol

Seite 2 - IST98202 R5 05/05

MAINTENANCEIn case of erroneous reading even after anaccurate calibration, the reference junction canbe contaminated or clogged.Pull out 2 mm (1/8&apo

Kommentare zu diesen Handbüchern