Seite 1 - Manganese High Range ISM

1Instruction ManualHI 96709CManganese High Range ISMwww.hannainst.com

Seite 2 - TABLE OF CONTENTS

10Under range: A blinking “0” indicates that the sample absorbsless light than the zero reference. Check the procedure and makesure you use the same c

Seite 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Seite 4 - ABBREVIATIONS

12• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor dirt. Wipe it thorough

Seite 5 - PRECISION AND ACCURACY

13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed

Seite 6 - PRINCIPLE OF OPERATION

14• Remove the cuvette.• Add the content of one packet ofHI 93709A-0 Manganese HR ReagentA, replace the cap and shake gently thecuvette for 2 minutes.

Seite 7 - FUNCTIONAL DESCRIPTION

15VALIDATION PROCEDURECalcium above 700 mg/LChloride above 70000 mg/LIron above 5 mg/LMagnesium above 100000 mg/LUse the validation procedure to ensur

Seite 8

16• After a few seconds, the display will show“-0.0-”. The meter is now zeroed and readyfor validation.• Remove the cuvette.• Place the CAL CHECKTM St

Seite 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Seite 10

18• After a few seconds the display will show“-0.0-”. The meter is now zeroed and readyfor calibration. The blinking “READ” asks forreading calibratio

Seite 11

19• Press GLP/ to edit the desired year(2000-2099). If the key is kept pressed,the year number is automaticallyincreased.• Press GLP/ to edit the desi

Seite 12 - MEASUREMENT PROCEDURE

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC

Seite 13

20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco

Seite 14

21• Press ZERO/CFM to restore the factorycalibration or press GLP/ again toabort factory calibration restore.• Press READ/ /TIMER to enter in thefact

Seite 15 - VALIDATION PROCEDURE

22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract

Seite 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Seite 17 - CALIBRATION PROCEDURE

24MAN96709 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Seite 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Seite 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96709 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Seite 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.0 to 20.0 mg/LResolution 0.1 mg/LAccuracy ±0.2 mg/L ±3% of reading @ 25°CTypical EMC Deviation ±0.1mg/LLi

Seite 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Seite 22 - ACCESORIES

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette h

Seite 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or t

Seite 24 - (ROMANIA)

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs