Hanna-instruments HI 96702 Bedienungsanleitung Seite 15

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 24
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 14
15
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK™ Standards, otherwise erroneous results will be obtained.
Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
Place the CAL CHECK™ Standard
HI 96702-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for validation.
Seitenansicht 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 23 24

Kommentare zu diesen Handbüchern

Keine Kommentare