METHODS
5.5.9.4 Sample Size
This option allows the user to enter the sample size. For External/Dissolution, enter the size
of the aliquot taken from the external vessel.
Before the titration is started, the user is asked again to enter the sample size. The sample
size (mass or volume) can be automatically acquired from the balance (when the balance
feature is enabled - see Section
4.11, Setup Balance Interface
)
5.5.9.5 Sample Density (
Sample Type: Volume
only)
Enter the density of the sample in g/mL. This allows calculation of mass-based result units
with volume-based sample sizes.