Hanna-instruments HI 901 Bedienungsanleitung Seite 58

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 148
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 57
5-18
METHODS
Threshold:
This parameter must be set by the user in according with the analysis.
The threshold represents the absolute value of the first derivative, expressed in |mV/mL|
(absolute value), below which the detection algorithm does not search for the equivalence
point.
Range is between 1 and 9999 mV/mL.
The recommended value for the threshold is around 40% of the estimated maximum absolute
value of the first derivative.
Depending on the titration curve profile, the following guiding ranges might be used:
TITRATION CURVE PROFILE THRESHOLD |mV/mL|
FLAT 1 to 450
NORMAL 450 to 1800
STEEP 1800 to 9999
Seitenansicht 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 147 148

Kommentare zu diesen Handbüchern

Keine Kommentare