Hanna-instruments HI 88703 Bedienungsanleitung Seite 13

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 32
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 12
13
Mark this position on the thicker white band on
the top of the cuvette with a water resistant
pencil.
Always use this position to align it with the
sign on the instrument top.
In order to match multiple cuvettes follow the next
steps:
Fill some cuvettes with high quality water
(<0.1NTU) up to the 10 mL mark.
Clean and oil the cuvettes as described before.
Turn the instrument ON.
Insert the first cuvette into the instrument and
press the Read key. Record the reading.
Mark this position on the thicker white band on
the top of the cuvette with a water resistant
pencil.
MATCHING MULTIPLE CUVETTES
Precise measurements require the use of a single cuvette. If it is not possible, cuvette selection
and matching must be performed before taking measurements.
Seitenansicht 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 31 32

Kommentare zu diesen Handbüchern

Keine Kommentare