The temperature has an effect on pH. The calibration buffer solutions are affected by temperature
changes to a lesser degree than normal solutions. During calibration the instrument will
automatically calibrate to the pH value corresponding to the measured or set temperature.
During calibration the instrument will display the pH buffer value at 25 °C.
pH BUFFER TEMPERATURE DEPENDENCE
PMETSREFFUBHp
C°F°10.403.810.01
023 10.484.823.01
514 00.444.842.01
0105 00.414.881.01
5195 00.473.821.01
0286 00.433.860.01
5277 10.403
.810.01
0368 20.472.869.9
5359 30.442.829.9
04401 40.412.888.9
TITRANT PREPARATION
- Take 10 mL HI 84442-50 titrant solution (accurate measured) and add to the clean 100
mL white bottle.
- Add 90 mL of distilled water.
- Mix carefully.
Use new obtained titrant on the further titrations.