VIEW/DELETE RECORDED pH DATA
To view or delete previously logged pH records press MENU key while in pH meter screen.
Press Recall to access the pH recall. A list of records is stored in the pH log.
Use the ARROW keys to scroll the list of records.
Press More to see detailed information about the highlighted record.
Press Delete to enter record deleting mode. Press Delete All to enter all records deleting mode.
If More is pressed a complete set of data is displayed.
Use ARROW keys when is displayed to scroll between the records.
If Delete was pressed the instrument will ask for confirmation.
Use the ARROW keys to focus on the record to be deleted.
Press Confirm to delete the record or ESC to return to the previous screen without deleting.
of measurements. If Manual Temperature Compensation (MTC) is desired, the temperature probe
must be disconnected from the instrument. The default temperature of 25 °C (77 °F) or the last
temperature reading will be displayed preceded by the symbol
and the “MTC” message.
The manually set temperature can now be adjusted with the ARROW keys (from -20.0 to 120.0 °C).