Hanna-instruments HI 84429 Bedienungsanleitung Seite 28

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 44
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 27
28
After the pump calibration is complete a confirmation message is displayed for a few seconds,
then the instrument will return to the titrator calibration menu and will display the new time
and date for the last pump calibration.
Notes: The calibration of the pump is independent of the selected range and concentration unit.
If an erroneous situation is encountered during the calibration, an error message is
displayed and the calibration can be restarted by pressing Restart.
If the calibration doesn’t complete within 6 minutes the error message “Too much
standard” will be displayed and the calibration can be restarted by pressing Restart
after a new sample is prepared.
Seitenansicht 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 43 44

Kommentare zu diesen Handbüchern

Keine Kommentare