Insert the cuvette into the instrument and press
the Read key. Record the reading.
Open the instrument lid, slightly rotate the
cuvette and take a new reading.
Repeat the last step until you read the lowest
NTU value.
Alternatively, keep the Read key pressed to
make continuous readings. After the first value
is displayed, open the lid and start rotating the
cuvette until the lowest NTU value is displayed.
Mark this position on the thicker white band on
the top of the cuvette with a water resistant
pencil.
Always use this position to align it with the
sign on the instrument top.
In order to match multiple cuvettes follow the next
steps:
Fill some cuvettes with high quality water
(<0.1NTU) up to the 10 mL mark.
Clean and oil the cuvettes as described before.
MATCHING MULTIPLE CUVETTES
Precise measurements require the use of a single cuvette. If it is not possible, cuvette selection
and matching must be performed before taking measurements.