45
COD MR
Select the
Oxygen Demand, (COD) MR
method following one of the procedures described in the
“Method Selection” section (see page 15).
Place the blank vial into the holder and push it completely down.
Press the Zero key. The instrument will perform a zero sequence and after a few seconds the display
will show “-0.0-”. Now the meter is zeroed and ready for measurement.
Remove the blank vial.
Place the sample vial into the holder and push it completely down.
Press Read.
The instrument will perform the reading.
The instrument displays concentration in mg/L of oxygen demand on the LCD.
INTERFERENCES
Interference may be caused by:
Chloride (Cl
-
) above 2000 mg/L.
Samples with higher chloride concentration should be diluted.