23
Bromine
BROMINE
SPECIFICATIONS
Range 0.00 to 8.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.08 mg/L ±3% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition
, DPD method. The reaction between bromine and the reagent causes a
pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93716-0 DPD Reagent 1 packet
REAGENT SETS
HI 93716-01 Reagents for 100 tests
HI 93716-03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
Select the
Bromine
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL