Hanna-instruments HI 83211 Bedienungsanleitung Seite 22

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 68
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 21
22
Ammonia LR
Remove the cuvette.
Add 4 drops of HI 93700A-0 First Reagent (6 drops for
seawater analysis). Replace the cap and mix the solution.
Add 4 drops of HI 93700B-0 Second Reagent (10 drops for
seawater analysis). Replace the cap and mix the solution.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and 30 seconds and
press Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of ammonia nitrogen
(NH
3
-N).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of ammonia (NH
3
) and ammonium (NH
4
+
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by: acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic
chloramines, sulfide, various aliphatic and aromatic amines.
Seitenansicht 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 67 68

Kommentare zu diesen Handbüchern

Keine Kommentare