Hanna-instruments HI 83208 Bedienungsanleitung Seite 68

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 67
68
Silver
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Fill a second cuvette up to the mark with 10 mL of the reacted sample.
Add 3 drops of Dispersing Agent (HI 93703-51),
replace the cap and invert gently to mix for about
10 seconds.
Insert the second cuvette into the instrument.
Press Read to start the reading. The instrument displays the results in mg/L of silver.
INTERFERENCES
Interference may be caused by:
Al
3+
above 30 mg/L Fe
2+
above 1.5 mg/L
Ca
2+
above 1000 mg/L as CaCO
3
Fe
3+
above 10 mg/L
Cd
2+
above 20 mg/L K
+
above 500 mg/L
Cl¯ above 8000 mg/L Mn
2+
above 25 mg/L
Co
2+
above 1.5 mg/L Mg
2+
above 1000 mg/L as CaCO
3
Cr
3+
above 20 mg/L Na
+
above 5000 mg/L
Cr
6+
above 40 mg/L Ni
2+
above 1.5 mg/L
Cu
2+
above 15 mg/L Pb
2+
above 20 mg/L
F¯ above 20 mg/L Zn
2+
above 30 mg/L
10 mL
# 2
# 2