1Instruction ManualHI 83208Multiparameter BenchPhotometerfor Water Conditioningwww.hannainst.comThis Instrument is inCompliance with the CE Directives

Seite 2 - TABLE OF CONTENTS

18 1910 mL• Remove the cuvette.• Add 4 drops of HI 93715A-0 First reagent (6 drops for seawateranalysis). Replace the cap and mix the solution.• Add 4

Seite 3 - PRECISION AND ACCURACY

20 21FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3%

Seite 4 - FUNCTIONAL DESCRIPTION

22 23• Reinsert the cuvette into the instrument.• Press READ to start the reading.• The instrument displays the results in mg/L of free chlorine.INTER

Seite 5 - USING CUVETTES

24 25Total Chlorine• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen

Seite 6 - HEALTH & SAFETY

26 27Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviat

Seite 7 - OPERATIONAL GUIDE

28 29Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±5% of readingTypical EMC ±1 μg/LDeviationLight Sou

Seite 8 - SETUPSETUP

30 31FluorideFLUORIDESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tungs

Seite 9 - HELP MODE

32 33Iron HRIRON HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±2% of readingTypical EMC ±0.01 mg/LDeviationL

Seite 10 - AMMONIA LOW RANGE

34 35Iron LRIRON LOW RANGESPECIFICATIONSRange 0 to 400 µg/LResolution 1 µg/LAccuracy ±10 µg/L ±8% of readingTypical EMC ±1 µg/LDeviationLight Source T

Seite 11 - FREE CHLORINE

36 37Iron LRINTERFERENCESInterference may be caused by:Cadmium above 4.0 mg/LChromium3+ above 0.25 mg/LChromium6+ above 1.2 mg/LCobalt above 0.05 mg/L

Seite 12 - INTERFERENCES

2 3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment,

Seite 13 - TOTAL CHLORINE

38 39Manganese HR• Add the content of one packet of HI 93709B-0 Sodium Periodatereagent. Replace the cap and shake gently until completely dissolved.•

Seite 14 - COPPER HIGH RANGE

40 41Manganese LR• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagentsolution to each cuvette, replace the caps and shake gently.• Add 1 mL of the

Seite 15 - COPPER LOW RANGE

42 43MolybdenumMOLYBDENUMSPECIFICATIONSRange 0.0 to 40.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight S

Seite 16 - FLUORIDE

44 45Molybdenum• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, waitfor five minutes and press READ.

Seite 17 - IRON HIGH RANGE

46 47Nickel HR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown priorto the measurement or, alternative

Seite 18 - 30"

48 49Nickel LR• Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, waitfor 15 minutes. Add one packet of H

Seite 19

50 51NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of readingTypical EMC ±0.1 mg/LDeviationLight Source Tungst

Seite 20 - MANGANESE HIGH RANGE

52 53Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of readingTypical EMC ± 0.1 mg/LDeviationLight Sour

Seite 21 - MANGANESE LOW RANGE

54 55pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pHTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow band int

Seite 22 - MOLYBDENUM

56 57Phosphate HRPHOSPHATE HIGH RANGESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±1 mg/L ±4% of readingTypical EMC Dev. ±0.1 mg/LL

Seite 23

4 5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wave

Seite 24 - NICKEL HIGH RANGE

58 59Phosphate LRPHOSPHATE LOW RANGESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of readingTypical EMC Dev. ±0.01

Seite 25 - NICKEL LOW RANGE

60 61PhosphorusPHOSPHORUSSPECIFICATIONSRange 0.0 to 15.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±4% of readingTypical EMC Dev. ±0.2 mg/LLight Sourc

Seite 26 - MEASUREMENT PROCEDURE

62 63SilicaSILICASPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviationLight Sourc

Seite 27 - DISSOLVED OXYGEN

64 65INTERFERENCESInterference may be caused by:Phosphate above 60 mg/LPhosphate above 75 mg/LSulfide and high concentration of ironEliminate color an

Seite 28

66 67Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker(the sample) and swirl gently to mix.• Press TIMER and the display w

Seite 29 - PHOSPHATE HIGH RANGE

68 69ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tu

Seite 30 - PHOSPHATE LOW RANGE

70 71The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messag

Seite 31 - PHOSPHORUS

72 73Description Range MethodAmmonia MR 0.00 to 10.00 mg/ L NesslerAmmonia LR 0.00 to 3.00 mg/ L NesslerChlorine, Free 0.00 to 2.50 mg/L DPDChlorine,

Seite 32

74 75WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose a

Seite 33

76Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699e-mail: [email protected]:Tel. (10) 88570068 • Fax (10) 88570060e-mail: hannachina@vi

Seite 34

6 7INSTRUMENT DESCRIPTION1) Cuvette Lid2) Liquid Crystal Display (LCD).3) Splash proof keypad.4) ON/OFF power switch5) Power input connector6) USB con

Seite 35 - REQUIRED REAGENT

8 9USING LIQUID AND POWDER REAGENTS• Proper use of the dropper:(a) for reproducible results, tap the dropper on the table for several times and wipe

Seite 36 - ERRORS AND WARNINGS

10 11Pay attention to push the cuvette completely down in the holder.• In order to avoid reagent leaking and to obtain more accurate measure-ments, cl

Seite 37 - ACCESSORIES

12 13POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note:

Seite 38 - HANNA LITERATURE

14 15Date / TimeThis option is used to set the instrument’s date and time.Press Modify functional key to change the date/time.Press the WX functional

Seite 39 - MAN83208 05/08

16 17Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source