Hydrazine
HYDRAZINE
SPECIFICATIONS
Range 0 to 400 μg/L
Resolution 1 μg/L
Accuracy ±4% of full scale at 25 °C
Typical EMC ±2 μg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, method
D1385-88,
p-Dimethylaminobenzaldehyde method. The reaction between hydrazine
and the liquid reagent causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93704-0 Liquid Reagent 24 drops
REAGENT SETS
HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Hydrazine
method using the procedure described in
the
Method Selection
section (see page 12).
Fill one cuvette up to the mark with 10 mL of distilled water.
Place the cap, insert the cuvette # 1 into the holder and
close the lid.
Press the Blank key to start adjusting the light level. The display will show “Blank done” when the
meter is ready to take a zero measurement.
Fill a second cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
10 mL
# 1
10 mL
# 2
# 1