1Instruction ManualHI 83203Multiparameter BenchPhotometerfor Aquaculturewww.hannainst.comThis Instrument is inCompliance with the CE Directives

Seite 2 - TABLE OF CONTENTS

18 1910 mlAmmonia LR• Remove the cuvette.• Add 4 drops of HI 93700A-0 First reagent (6 drops for seawateranalysis). Replace the cap and mix the soluti

Seite 3 - PRECISION AND ACCURACY

20 21Free Chlorine• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagent

Seite 4 - FUNCTIONAL DESCRIPTION

22 23Total ChlorineTOTAL CHLORINESPECIFICATIONSRange 0.00 to 3.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 m

Seite 5 - USING CUVETTES

24 25• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, waitfor 2 minutes and 30 seconds and press READ

Seite 6 - HEALTH & SAFETY

26 27Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate. Replacethe cap and shake gently for about 15 seconds.

Seite 7 - SETUPSETUP

28 29Copper LR• Remove the cuvette.• Add the content of one packet of HI 93747-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.

Seite 8 - HELP MODE

30 31Nitrate• Replace the cap and immediately shake vigorously upand down for exactly 10 seconds. Continue to mix byinverting the cuvette gently for 5

Seite 9 - AMMONIA MEDIUM RANGE

32 33Nitrite HR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown priorto the measurement or, alternativ

Seite 10 - AMMONIA LOW RANGE

34 35• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown priorto the measurement or, alternatively, wait

Seite 11 - FREE CHLORINE

36 37Dissolved Oxygen• Let the sample stand and the flocculent agent will start to settle.• After approximately 2 minutes, when the upper half of the

Seite 12 - TOTAL CHLORINE

2 3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment,

Seite 13 - INTERFERENCES

38 39pH• Reinsert the cuvette into the instrument.• Press the READ key to start the reading. The instrument displaysthe pH value.pHpHSPECIFICATIONSRan

Seite 14 - COPPER HIGH RANGE

40 41Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until completely

Seite 15 - COPPER LOW RANGE

42 43Phosphate LR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, alternat

Seite 16 - SPECIFICATIONS

44 45Description Range MethodAmmonia MR 0.00 to 10.00 mg /L NesslerAmmonia LR 0.00 to 3.00 mg/ L NesslerChlorine, Free 0.00 to 2.50 mg/ L DPDChlorine,

Seite 17 - NITRITE HIGH RANGE

46 47WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose a

Seite 18 - NITRITE LOW RANGE

48Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699e-mail: [email protected]:Tel. (10) 88570068 • Fax (10) 88570060e-mail: hannachina@vi

Seite 19 - DISSOLVED OXYGEN

4 5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wave

Seite 20

6 7INSTRUMENT DESCRIPTION1) Cuvette Lid2) Liquid Crystal Display (LCD).3) Splash proof keypad.4) ON/OFF power switch5) Power input connector6) USB con

Seite 21 - PHOSPHATE HIGH RANGE

8 9(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hangingon the tip of the syringe, if so eliminate them

Seite 22 - PHOSPHATE LOW RANGE

10 11POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note:

Seite 23 - ERRORS AND WARNINGS

12 13In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are general parameters that

Seite 24 - ACCESSORIES

14 15HI 83203 offers an interactive contextual help mode that assists the user at any time.To access help screens press HELP.The instrument will displ

Seite 25 - MAN83203 05/08

16 17• Remove the cuvette.• Add 4 drops of HI 93715A-0 First reagent (6 drops for seawateranalysis). Replace the cap and mix the solution.• Add 4 drop