Press Timer and the display will show the countdown prior to zeroing the blank or, alternatively, wait
for two minutes and press Zero. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Insert the other cuvette (# 2) with the reacted sample into the
instrument.
Press Read to start reading. The instrument displays the results in mg/L of fluoride.
Note: For wastewater or seawater samples, before performing measurements, distillation is required.
For most accurate results, use two graduated pipettes to deliver exactly 8 mL of distilled water and
8 mL of sample.
Fluoride
INTERFERENCES
Interferences may be caused by:
Alkalinity (as CaCO
3
) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Chloride above 700 mg/L
Phosphate, ortho above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.
Fluoride
FLUORIDE
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition
, SPADNS method. The reaction between fluoride and the liquid reagent
causes a red tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93729-0 SPADNS Reagent 4 mL
REAGENT SETS
HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 124.
MEASUREMENT PROCEDURE
Select the
Fluoride
method using the procedure described in
the
Method Selection
section (see page 12).
Add 2 mL of HI 93729-0 SPADNS Reagent to two cuvettes.
Fill one of the cuvettes with distilled water up to the mark,
replace the cap and invert several times to mix.
Fill the other cuvette with sample up to the mark, replace the
cap and invert several times to mix.
Place the cuvette with the reacted distilled water (# 1) into
the holder and close the lid.
# 2
# 1
# 1
# 2
10 mL10 mL
2 mL