Setup Item Description Valid Values Default
Analog Output 1
O.13 Analog output 1 0 to f.s. 1999 μS/cm
maximum value (O_VARMIN1 ≤
(O_VARMAX1) O_VARMAX1 -5% f.s. for
O_VARMAX1, O_VARMIN1
≤ O_HOLD1 ≤ O_VARMAX1)
O.14 Analog output 1 User selected value, Previous value
hold mode (note 13) Previous value
O.15 Analog output 1 0 to f.s. 1000 μS/cm
hold mode value (it must be within the
(O_HOLD1) O_VARMIN1 to
O_VARMAX1 interval)
Analog Output 2
O.20 Analog output 2 Recorder (temperature), Recorder
mode (notes 1, 17) Control-setpoint 2
(conductivity/concentration
/temperature)
O.21 Analog output 2 0-20 mA or 4-20 mA
type 4-20 mA
O.22 Analog output 2 -30.0 to 130.0°C 0.0°C
minimum value (O_VARMIN2 ≤
(O_VARMIN2) O_VARMAX2 -10°C,
O_VARMIN2 ≤ O_HOLD2
≤ O_VARMAX2)
O.23 Analog output 2 -30.0 to 130.0°C 100.0°C
maximum value (O_VARMIN2 ≤
(O_VARMAX2) O_VARMAX2 -10°C,
O_VARMIN2 ≤ O_HOLD2
≤ O_VARMAX2)
O.24 Analog output 2 User selected value, Previous value
hold mode (note 13) Previous value
O.25 Analog output 2 -30.0 to 130.0°C 25.0°C
hold mode value (it must be within the
(O_HOLD2) O_VARMIN2 to
O_VARMAX2 interval)
Setup Item Description Valid Values Default
Other Parameters
C.70 Hold mode end delay 00 to 99 seconds 00 sec
C.80 On/Off control 00:00 to 30:00 min. 00:00 min
action delay (note 26)
OUTPUT
Relays
O.01 Relay 1 mode Disabled, Control-
(see notes 16, 17) Control-setpoint 1, setpoint 1
Control-setpoint 2,
Simple cleaning,
Hold mode
O.02 Relay 2 mode Disabled, Control-
(see notes 16, 17) Control-setpoint 1, setpoint 2
Control-setpoint 2,
Simple cleaning,
Hold mode
O.03 Relay 3 mode Disabled, Disabled
(see notes 12, 16) Simple cleaning,
Advanced cleaning,
Hold mode
O.04 Relay 4 mode Disabled, Disabled
(see notes 12, 16) Simple cleaning,
Advanced cleaning,
Hold mode
O.05 Hold digital output Disabled or Hold mode
Hold mode
Analog Output 1
O.10 Analog output 1 Recorder or Recorder
mode (notes 1, 17) Control-setpoint 1
O.11 Analog output 1 0-20 mA or 4-20 mA
type 4-20 mA
O.12 Analog output 1 0 to f.s. 0 μS/cm
minimum value (O_VARMIN1 ≤
(O_VARMIN1) O_VARMAX1 -5% f.s. for
O_VARMAX1, O_VARMIN1
≤ O_HOLD1 ≤ O_VARMAX1)