HI 4008 Cupric Half-cellHI 4008 Cupric Half-cell
HI 4008 Cupric Half-cellHI 4008 Cupric Half-cell
HI 4008 Cupric Half-cell
HI 4108 Cupric Combination ElectrodeHI 4108 Cupric Combination Electrode
HI 4108 Cupric Combination ElectrodeHI 4108 Cupric Combination Electrode
HI 4108 Cupric Combination Electrode
I. I.
I. I.
I.
Introduction:Introduction:
Introduction:Introduction:
Introduction:
The Hanna HI 4008 and HI 4108 are ion selective elec-
trodes designed for the measurement of cupric ions in
aqueous solutions. The HI 4008 is a solid state half-cell
sensor that requires a separate reference. The HI 4108 is a
combination ion selective electrode.
IIII
IIII
II.
SpecificationsSpecifications
SpecificationsSpecifications
Specifications
Type: Solid State electrode with
a cupric sulfide /silver sul-
fide membrane.
Interfering ions: Silver and mercury must
be absent. Any ion that forms a more insoluble sulfide than
cupric sulfide will interfere with the electrode response. These
include bismith (Bi
3+
), iron (Fe
2+
) ,bromide(Br
-
) and
chloride (Cl
-
) if they are in high concentrations. Long term
exposure in this ion will poison the cupric surface. Oxidation
of the cupric surface will also poison the electrode.
Operating Temperature: 0-80°C
Operating pH: 3 to 7 pH
(4-6 recommended)
Dimensions: 12 mm (OD) X 120 mm
nominal insertion
(0.47X 4.72)
Connection: BNC
Ion(s) measured: Cupric (Cu
2
++
++
+
)
Measurement range: 0.1
M to 1X 10
-6
M
6355 to 0.065 ppm