1Instruction ManualHI 95762CFree Chlorine Meterfor Drinking Water Applicationwww.hannainst.comThis Instrument is inCompliance with the CE Directives20
10The flashing digits indicate that:1) the month can be set2) the day can be set“Storage”: this prompt appears for 1 second at the end of thecalibrati
2PRELIMINARY EXAMINATION... 3GENERAL DESCRIPTION ...
3Please examine this product carefully. Make sure that the instrument is not damaged. If anydamage occured during shipment, please notify your Dealer.
4The HI 95762 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical instru
5Range 0.000 to 0.500 mg/LResolution 0.001 mg/LPrecision ±0.004 mg/L @ 0.200 mg/LTypical EMC Deviation ±0.001 mg/LLight Source Tungsten lamp with narr
6Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam crosses a substance, some o
7A microprocessor controlled special tungsten lamp emits radiation which is first optically conditionedand beamed to the sample contained in the cuvet
8KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO: to zero the meter prior to measurement.• READ/TIMER: this is a bi-functional key, ju
9This prompt appears for 1 second each time the instrument isturned on.This indicates that the meter is in a ready state and zeroing can beperformed.S