Wipe the cuvette thor-
oughly with a lint-free
tissue (HI 93703-70)
before inserting into
the measurement
cell. The cuvette
must be completely
free of fingerprints
and other oil or dirt,
particularly in the
area where the light
goes through (ap-
proximately the bot-
tom 2 cm/1 inch of
the cuvette).
Place the cuvette
into the cell and
check that the notch
on the cap is posi-
tioned securely into
the groove.
The mark on the
cuvette cap should
point towards the
LCD.
Press the READ/
key and the LCD will
display a blinking
“SIP” (Sampling in
Process). The tur-
bidity value will ap-
pear after approxi-
mately 25 seconds.
OPERATIONAL GUIDE
To prepare the instrument for taking
measurements, first install the batteries (see
Battery Replacement section on page 17) and
then turn the instrument on.
To maximize the battery life the meter is
automatically switched off after 5 minutes of
non-use. To reactivate it, simply press the ON/
OFF key.
MEASUREMENT PROCEDURE
Turn the meter on by
pressing the ON/OFF
key.
The meter will carry
out a self-test display-
ing a full set of figures.
After the test, the LCD
will change to the mea-
surement mode.
When the LCD dis-
plays “----” the meter
is ready to measure.
Fill a clean cuvette up
to one quarter inch
(0.5 cm) from its rim
with the thoroughly
agitated sample.
Allow sufficient time for bubbles to escape
before securing the cap.