1pH/mV/°CBench Meterswith Custom BuffersHI 110 HI 111HI 112 HI 113Instruction ManualpHMRMEMSETCCRANGEGLPCFMCALpH / ORP MeterHI 113Select Setup Ite
10RELATIVE mV MEASUREMENTS (HI 111 and HI 113) Press RANGE until rEL message will be displayed on the secondary LCD linefor about one second and m
11PREPARATIONPour small quantities of the buffer solutions into clean beakers. If possible, useplastic or glass beakers to minimize any EMC interferen
12THREE- (HI 110, HI 111) / FIVE-POINT (HI 112,HI 113) CALIBRATION Immerse the pH electrode and the temperature probeapproximately 4 cm (1½) into a
13 If necessary, press the ARROW keys to select a different buffer value. The tag will blink on the LCD until the reading isstable. When the re
14 To clear calibration parameters for all uncalibrated buffers startingwith current buffer, press and hold down CFM, then press CAL. Ifthis procedur
15GOOD LABORATORY PRACTICE (GLP)GLP is a set of functions that allows storage and retrieval of data regarding themaintenance and status of the electro
16 The pH calibration buffers in calibrating order.The first pH calibration buffer:The second pH calibration buffer:The third pH calibration buffer:N
17Setup mode allows viewing and modifying the following parameters: Calibration Alarm Time-Out Custom buffer 1 Custom buffer 2 One-point calibrati
18CALIBRATION ALARM TIME-OUT SETPress CAL when the calibration time-out is displayed. Calibration time-out(OFF or 1 to 7 days) will start blinking.Pre
19INSTRUMENT ID PARAMETER SET(HI 112, HI 113 only)Press CAL when the instrument ID is displayed. The instrument ID (0000 to9999) will start blinking.P
2WARRANTYHI 110, HI 111, HI 112 and HI 113 are guaranteed for two years againstdefects in workmanship and materials when used for their intended purpo
20Press RANGE. The day will start blinking.Press the ARROW keys to change the day.Press CFM to save the modified value.Press CAL to escape without sav
21All the instruments are factory calibrated for temperature.Hannas temperature probes are interchangeable and no temperaturecalibration is needed wh
22 Use the ARROW keys to set the reading on the secondary LCD to that ofthe hot water. When the reading is stable and close to the selectedcalibrati
23 mV CALIBRATION(for technical personnel only)All the instruments are factory calibrated for mV.Hannas ORP electrodes are interchangeable and no mV
24RELATIVE mV CALIBRATION (HI 111AND HI 113) Press CAL when the instrument is in RELATIVE mV measurement mode.The mV and tags will blink. Absol
25Data transmission from the instrument to the PC can be done with the HI92000 Windows® compatible software (optional). HI 92000 also offersgraphing a
26CHR n Change the instrument range according with the parametervalue (n): n=0 pH range/0.001 resolution n=1 pH range/0.01 resolution n=3 mV range
27The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du
28PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal withelectrodes. Th
29For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1) below the fill hole,add HI 7082 or HI 8082 3.5M KCl Electrol
3PRELIMINARY EXAMINATIONRemove the instrument from the packing material and examine it carefully tomake sure that no damage has occurred during shippi
30CLEANING PROCEDURE General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour. Protein Soak in Hanna HI 7073 or HI
31TROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnoperwolS.tfird.edortceleHpytriD nipitedortceleehtkaoS1608IHro1607IHsetunim03rofnoituloseh
32Typical Electrode LifeAmbient Temperature 1 3 years90 ºC Less than 4 months120 ºC Less than 1 monthAlkaline ErrorHigh concentrations of sodium ion
33pH BUFFER SOLUTIONSHI70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI70010P pH 10.01 Buffer Sachets,
34ORP PRETREATMENT SOLUTIONSHI7091L Reducing Pretreatment Solution, 460 mLHI7092L Oxidizing Pretreatment Solution, 460 mLpH ELECTRODESAll electrodes
35HI1330BGlass-body, semimicro, single junction, refillable, combination pH electrode.Use: laboratory, vials.HI1331BGlass-body, semimicro, single ju
36120 mm4.7"12 mm0.5"FC 10075 mm2.95"6 mm0.25"FC 200120 mm4.7"12 mm0.5"FC 210120 mm4.7"12 mm0.5"9.5mm DIA0.37&
37Kynar® is registered Trademark of Pennwalt Corp.Ultem® is registered Trademark of General Electric Co.HI1413BGlass-body, single junction, flat
38EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap
39RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th
41) Liquid Crystal Display (LCD).2) CFM/GLP key, to confirm different values or to display Good LaboratoryPractice information.3) CAL key, to enter an
40SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666 Fax (03) 9769.0699China:Tel. (10) 88570068 Fax (10) 88570060Egypt:Tel. & F
5HI 110 AND HI 111SPECIFICATIONSEGNARHp00.61ot00.2)ylno111IH,PRO&ESI(Vm9.996±)ylno111IH,PRO&ESI(Vm0002±Cº0.021ot9.9NOITULOSERHp10.0)ylno111I
61) Liquid Crystal Display (LCD).2) CFM/GLP key, to confirm different values or to display Good LaboratoryPractice information.3) CAL key, to enter an
7HI 112 AND HI 113SPECIFICATIONSEGNARHp00.61ot00.2Hp000.61ot000.2)ylno311IH,PRO&ESI(Vm9.999±)ylno311IH,PRO&ESI(Vm0002±Cº0.021ot9.9NOITULOSE
8OPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the power supply socket.Note: These instruments use non volatile memory to retain the p
9If measurements are taken successively in different samples, it is recommendedto rinse the electrode thoroughly with deionized water or tap water and