1Instruction ManualHI 8014 • HI 8915HI 931000Portable pH Meterswww.hannainst.com

Seite 2 - GENERAL DESCRIPTION

18 19For manual temperature compensation during calibration, pleaserefer to the following chart.pH VALUES AT VARIOUSTEMPERATURESFor instance, if the b

Seite 3 - SPECIFICATIONS HI8014

20 21If the bulb and/or junction are dry, soak the electrode in HI 70300Storage Solution for at least one hour.For refillable electrodes:If the refill

Seite 4 - SPECIFICATIONS HI8915

22 23Inorganic Soak in Hanna HI 7074 Inorganic Cleaning Solu-tion for 15 minutes.Oil/grease Rinse with Hanna HI7077 Oil and Fat CleaningSolution.IMPOR

Seite 5 - SPECIFICATIONS HI931000

24 25ACCESSORIESpH CALIBRATION SOLUTIONSHI 70004P pH 4.01 buffer solution, 20 mL sachet, 25 pcsHI 7004M pH 4.01 buffer solution, 230 mL bottleHI 7004L

Seite 6 - OPERATIONAL GUIDE

26 27FC210B Glass body, double junction, conic, Viscolene, non-refillable combination pH electrode. Use: milk, yogurt.FC220B Glass body, triple cerami

Seite 7 - AFTER MEASUREMENTS

28MANPHR4 07/05SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666 • Fax (03) 9769.0699China:Tel. (10) 88570068 • Fax (10) 88570060Eg

Seite 8 - PROCEDURE FOR HI 8915

23Dear Customer,Thank you for choosing a Hanna product. Please read this instructionmanual carefully before using the meter. This manual will provide

Seite 9 - PROCEDURE FOR HI 931000

45FUNCTIONAL DESCRIPTION HI8014ins rum nspHpH meterHI 8014°C mV/pHAPpH°COFFmV21341) BNC socket for combination pH electrodes.2) Liquid Crystal Display

Seite 10 - HI 8014 and HI 8915

67FUNCTIONAL DESCRIPTION HI8915pHONOFFHOLDmV°CpHHI 8915ATCpHmeter+-mV OUTPUTSLOPE4.0110.01OFFSET6.867.00CALIBRATION1729410511836121) BNC socket for co

Seite 11 - PREPARATION

89FUNCTIONAL DESCRIPTION HI931000ins rum nsCFM°CCALOFFONHI 9310001234561) DIN socket for combination pH/°C electrode.2) Liquid Crystal Display (LCD).3

Seite 12 - TROUBLESHOOTING

10 11OPERATIONAL GUIDEINITIAL PREPARATIONEach meter is supplied complete with a 9V battery. Slide off thebattery compartment cover on the back of the

Seite 13 - S = SCREW-TYPE CONNECTOR

12 13Notes:• In order to take accurate pH measurements, make sure that theinstrument is calibrated for pH before use.• For a faster response time when

Seite 14 - OTHER ACCESSORIES

14 15pH CALIBRATIONFor best accuracy, frequent calibration of the instrument is recom-mended. The instrument should be recalibrated for pH:a) Whenever

Seite 15 - MANPHR4 07/05

16 17• Stir gently and wait a couple of minutes for thermal equilibrium.Note: The electrode should be submerged approximately 4 cm (1½")into the